Üldinfo
Metoodika XRF
Lisainfo The bulk chemical composition and Fe2O3 total of the rocks was determined by XRF spectrometry in the laboratories of the All-Russian Geological Institute (VSEGEI), St. Petersburg. The samples were crushed by a ball mill till 5 microns. The weight of samples for analysis was 3 grams. The sample was pressed to stainless steal cuvette. The pressure was 1 ton. For the measurements the X-ray spectrometer SPM-25 (1988) was used. As the source of primary X-ray radiation there was used the X-ray tube with rhodium anode (3RXB). The time of exposition was 80 seconds. The reliability of the results of X-ray fluorescence anal¬ysis was controlled by standards that the GOST (USSR state standards) certificates have. The analyses were made according to the Instruction NSAM N-313-RS (VIMS, VSEGEI, 1990). The precision of the measurements was regulated by Instruction (OSM 41-08-212-82). All concentrations below the detection level were set to one-half of the detection limit.
Lokaliteet
Sügavus 306.8
Stratigraafia
Asutus TalTech geoloogia instituut
Kirje lisatud 2013-10-28
Kirje muudetud 2014-01-09
Tulemused
ParameeterVäärtusTekstViga
SiO2 [%]93.92
TiO2 [%]0.165
Al2O3 [%]0.402
Fe2O3total [%]0.471
MnO [%]<.03000
MgO [%]0.576
CaO [%]1.81
Na2O [%]<.05000
K2O [%]<.05000
P2O5 [%]1.08
S [%]0.0635
Ba [ppm]<.01000
Cr [ppm]<.00500
V [ppm]<.01000
LOI [%]1.31